HARDWARE OPTIONS SCREEN:
From the Service Mode screen, press the Options screen button then the Hardware
Options button.
ONLINE PHT: This option is used to enable or disable the online pressure
Line M:
holding test option. This setting will toggle from Yes to No by pressing the
Online PHT button. Once the option is set press the [CONFIRM] key and
the X will change from gray to blue. Select Yes or No as dictated by
facility management. Record this setting on Installation Checklist.
ART CHAMBER: This option is used to select for a prepump or a
Line N:
postpump arterial chamber tubing configuration. This setting will toggle
from Pre to Post by pressing the Art Chamber button. Once the option is
set press the [CONFIRM] key and the X will change from gray to blue.
Select Pre or Post as dictated by facility management. Note: Dipswitch #2
on the blood pump must agree with this setting. Record this setting on
Installation Checklist.
Line O:
MAX UF RATE: This option is used to set the maximum allowable
ultrafiltration rate selectable by the user. To change this setting, select the
Max UF Rate data button then use the [
desired volume from 1000, 2000, 3000, or 4000ml/hr. When the desired
amount has been reached, press [CONFIRM] key. Record this setting on
the Installation Checklist.
Line P:
AUDIBLE ALARM: This option is used to enable or disable audible alarm
when the lines are off the shunt. This setting will toggle from Yes to No by
pressing the Audible Alarm button. If Yes is selected, audible alarms will
be heard when the lines are off the shunt. If No is selected, no audible
alarms will be heard when the lines are off the shunt and no blood is sensed.
NOTE: If blood is sensed, audible alarms will be heard regardless of the
Audible Alarm option setting.
Once the option is set press the [CONFIRM] key and the X will change
from gray to blue. Select Yes or No as dictated by facility management.
Record this setting on Installation Checklist
T&C MODE: This option is not used during the machine installation process
Line Q:
and must always have NO selected before the machine is released for use.
Line R:
0 ART LIMIT: This option is used to select if the upper arterial alarm limit
is moved to –15mmHg when blood is sensed. This setting will toggle from
Yes to No by pressing the 0 Art Limit button. If Yes is selected, the upper
arterial alarm limit will move to –15mmHg when blood is sensed. If No is
selected, the upper arterial alarm limit will NOT move to –15mmHg when
blood is sensed. Once the option is set press the [CONFIRM] key and the X
will change from gray to blue. Select Yes or No as dictated by facility
management. Record this setting on Installation Checklist.
P/N 507300 Rev. A
Page
5 of 17
=
>