2008K INSTALLATION CHECKLIST INSTRUCTIONS
Written By:
J. Fox
Distribution: Nat'l Svc Mgr (M. Lowery), TS Mgr (K. Cogswell), QSRA (M. Owen), QE (R. Fox), Eng (D.Gavin), Tng (B. Hassard)
PURPOSE:
The purpose of these instructions is to outline the process of installing a 2008K Hemodialysis
machine.
EQUIPMENT NEEDED:
Digital Volt Meter, Electrostatic Kit, Temperature/Conductivity meter and hydraulic pressure
gauges. WARNING: Use recommended ESD procedures when handling electronic devices.
PROCEDURES:
Use form P/N 490051 INSTALLATION CHECKLIST. Complete the upper information section
completely, this includes unit name and address, machine serial number, all software versions,
shipping damage and record any discrepancies with the shipped accessories.
SET THE K BLOOD PUMP DIPSWITCHES: With the machine power OFF, locate the
Line 1
dipswitches on the LP955 blood pump board. Set the dipswitches uses Table 1 as reference.
Dip
Switch
1
2
3
4
5
6
7
8
RINSE MACHINE WITH HEATER OFF: Insert the acid and bicarb connectors into
Line 2
the ports on front of machine and place dialysate connectors into the shunt. Power the
machine ON and wait for the startup screen to appear. Verify that the heater circuit
breaker on the back of the power supply is in the OFF position. On the startup screen,
press the Rinse button. Allow the machine to complete the rinse cycle (minimum 30
minutes). Record the rinse completion on the Installation Checklist.
Reviewed By:
R. Hassard
Description
ON for future use
OFF for use on the 2008K machine.
ON for Pre-Pump Arterial Pressure displays (-280 to +300 mmHg)
OFF for Post-Pump Arterial Pressure display (-80 to +500 mmHg)
Not Used
ON for Blood pump stop alarm after 30 seconds.
OFF for Blood pump stop alarm after 15 seconds.
Not Used
Not Used
Not Used
ON for tubing diameter selections 2 to 10mm (0.2mm increments)
OFF for tubing diameter selections 2.6, 4.8, 6.4 and 8.0mm only
Table 1
P/N 507300 Rev. A
Page
Approved By:
ECO# 7131
1 of 17
Effective Date: