Fresenius Medical Care 2008k Technical Manual page 5

Hemodialysis machine
Hide thumbs Also See for 2008k:
Table of Contents

Advertisement

Section I – Specifications
Water
Back Flow Prevention
Water Pressure
Water Temperature
Quality
Drain
Rinsing
Concentrate Supply
Concentrate Pressure
Dialysate Flow
Adjustment Range
Partial Dialysate Collection
Proportional Mixing System
Acid
Acetate
Adjustment Range
Bicarbonate
I-2
Integral back flow prevention provided by external vent
to atmosphere in water inlet circuit.
Min 20 psi; max 105 psi
Min 10 °C; max 25 °C
AAMI Standard, RD5, 1992
3 feet maximum height. Must comply with local codes
and must maintain a free fall air gap between drain hose
and building drain.
Temperature 37° C. Flow rate 620 ml/min. Time between
10 and 60 minutes (internally selectable)
Max suction height 3 feet
Max supplied pressure 2 psi
Sequential (0)/100/200/300/400/500/600/700/800
ml/ min., selectable in the Dialysis screen;
Dialysate on/off key
From Drain line, intermittent collection using a 3 Liter
PD drain bag as a collection device with a Safe-Lock
connector (optional).
Volumetric, selectable:
1:34
1:44
1:35.83
1:35.1
1:34
+
130 to 155 mEq/L Na
Volumetric, selected with associated acid ratio
1:27.6
1:19.1
1:24.5
1:28.6
2008K Technician's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents