Fresenius Medical Care 2008k Technical Manual page 140

Hemodialysis machine
Hide thumbs Also See for 2008k:
Table of Contents

Advertisement

Section VII – Troubleshooting
SCREEN 4: ELECTRONICS
LEAK =
Blood leak receiver (in Vdc). Sensor cable connector 2 pin 6. Sensor board ADC input
channel 0.
DIMN =
Blood leak dimness receiver (in Vdc). Sensor cable connector 2 pin 7. Sensor board
ADC input channel 1.
PDIA =
Dialysate pressure (in Vdc). Sensor cable connector 2 pin 8 and 10. Sensor board
ADC input channel 4.
TEMP =
Temperature monitor NTC3 (in Vdc). Sensor cable connector 2 pin 12. Sensor board
ADC input channel 5.
ABPR =
Arterial blood pump rate (in Vdc). Sensor board input pin A28. Sensor board ADC
input channel 6.
PART =
Arterial pressure (in Vdc). Sensor board input pin A27. Sensor board ADC input
channel 2.
5V =
5V power supply (in Vdc). Sensor board input pin C31. Sensor board ADC input
channel 8.
PVEN =
Venous pressure (in Vdc). Sensor board input pin A32. Sensor board ADC input
channel 3.
24V =
24V power supply (in Vdc). Sensor board input pin C30. Sensor board ADC input
channel 9.
ASET =
Arterial blood pump set (in Vdc). Sensor board input pin A12. Sensor board ADC
input channel 10.
VSET =
Venous (single needle) blood pump set (in Vdc). Sensor board input pin A11. Sensor
board ADC input channel 11.
VBPR =
Venous (single needle) blood pump rate (in vdc). Sensor board input pin A29. Sensor
board ADC input channel 7.
XTRM =
Last byte received from Transterm (actuator board).
CMS =
Undefined.
The following signals are Self-Test signals:
RAWT
MODE#
DOING
TSTT
CURT
DELAY
The following are functional board communication data used in software development:
XTMI
FDS#
FDSX
BPMI
BPMO
00F2
00F4
FDSO
VII-8
GOTT
TIME
FAIL#
XTMD
FDSR
FDSI
00F6
2008K Technician's Manual

Hide quick links:

Advertisement

Table of Contents
loading