EXT PRE-RINSE: This option is used to enable or disable the extended
Line E:
pre-rinse function. This setting will toggle from Yes to No by pressing the
Ext Pre-Rinse button. If this option is set to Yes, the hot rinse pre-rinse is
changed from 7 minutes to 20 minutes. Once the option is set press the
[CONFIRM] key and the X will change from gray to blue. Select Yes or
No as dictated by facility management. Record this setting on Installation
Checklist.
ALARM AT SVS END: This option is used to enable or disable an alarm
Line F:
at the completion of an SVS program. This setting will toggle from Yes to
No by pressing the Alarm at SVS end button. Once the option is set press
the [CONFIRM] key and the X will change from gray to blue. Select Yes
or No as dictated by facility management. Record this setting on
Installation Checklist.
ARTERIAL WIDTH: This option is used to select the type of arterial
Line G:
pressure alarm limit window. By pressing the Arterial Width button, a
selection menu appears. Used the [
following choices as dictated by facility management:
User Selectable
120
160
200
Line H:
VENOUS WIDTH: This option is used to select the type of venous
pressure alarm limit window. By pressing the Venous Width button, a
selection menu appears. Used the [
following choices as dictated by facility management:
User Selectable
100 Asymetric
120
160
200
=
] and [
The width of the arterial alarm window is not fixed and
can be changed by the user on the Test & Options
screen during dialysis.
Arterial alarm window fixed at a width of 120mmHg.
Arterial alarm window fixed at a width of 160mmHg.
Arterial alarm window fixed at a width of 200mmHg.
=
] and [
The width of the venous alarm window is not fixed and
can be changed by the user on the Test & Options
screen during dialysis.
Venous alarm window fixed at a width of 100mmHg
and the lower limit is at 25mmHg below the actual
value after the pressure settles within 1 minute.
Venous alarm window fixed at a width of 120mmHg.
Venous alarm window fixed at a width of 160mmHg.
Venous alarm window fixed at a width of 200mmHg.