AFTER 24 STROKES THE
BURET SHOULD INDICATE
BETWEEN 0.90 AND 1.10 ml.
SET THE FLUID LEVEL AT
EXACTLY 25 ml TO BEGIN.
LINE TO THE FLUID SAMPLE/ULTRAFILTRATE
OUTPUT PORT.
Select the Target data button.
5.
Set the Target value to 24 strokes, then press the [CONFIRM] key.
Press the [CONFIRM] key again. The UF pump will stroke and the screen
6.
will change.
Wait until the Target value reaches zero.
7.
Page 22
Figure 4. UF Pump Volume Measurement.
Fresenius 2008K Calibration Procedures
P/N 507296 Rev. A
ml
0
1
25