Fresenius Medical Care 2008k Technical Manual page 74

Hemodialysis machine
Hide thumbs Also See for 2008k:
Table of Contents

Advertisement

From the Calibrate Sensors screen, select the Arterial Pressure screen button.
The screen will change to the following:
N
Note: Depending on the HARDWARE OPTION set for the ART PUMP
option, the above display may read POSTPUMP instead of PREPUMP.
The bargraph scale will also be different (500 to –80mmHg).
Open arterial transducer port P
1.
(atmospheric pressure).
N
Note: Use the Arterial Level Adjust [
panel to abort the calibration process without changing the stored DAC
value on the blood pump.
Press the [Start/Stop] key on the blood pump and its display will change to
the following:
Press the [Start/Stop] key again and the small zero will start to flash.
Fresenius 2008K Calibration Procedures
on the blood pump module to air
ART
.
=
] key on the blood pump control
P/N 507296 Rev. A
Page 27