Section VII – Troubleshooting
SCREEN 1: HYDRAULICS (Cont.)
FLWP =
255 minus this value = DAC value sent to flow pump from the actuator board.
PHT ACT = On-line pressure holding test is active. (1 = yes and 0 = no).
DEAP =
255 minus this value = DAC value sent to deaeration pump from the actuator board.
NPHT =
Seconds until next on-line pressure holding test.
TEMP =
Calculated temperature (example 370 = 37.0°C).
PHT0 =
Result of PHT first valve combination.
5V EST. =
Estimated 5V from the power supply.
PHT1 =
Results of PHT second valve combination.
12V EST. =
Estimated 12V from the power supply
BPCE =
Blood pump communication errors. This is a number of how many errors have
occurred.
SALV =
Saline volume. Not used.
VERR =
Valve error (1 = yes : 0 = no). Not used.
VII-4
2008K Technician's Manual