ACID ALERT DEFAULT: This option is used to set the acid alert default
Line I:
volume. To change this setting, select the Acid Alert Default data button
then use the [
liters. When the desired amount has been reached, press [CONFIRM] key.
Record this setting on the Installation Checklist.
BIC ALERT DEFAULT: This option is used to set the bicarbonate alert
Line J:
default volume. To change this setting, select the Bic Alert Default data
button then use the [
9.9 liters. When the desired amount has been reached, press [CONFIRM]
key. Record this setting on the Installation Checklist.
PRIME AMOUNT: This option is used to set the priming volume used
Line K:
during prime. To change this setting, select the Prime Amount data button
then use the [
When the desired amount has been reached, press [CONFIRM] key. Record
this setting on the Installation Checklist.
RECIRCULATE: This option is used to enable or disable the recirculation
Line L:
function. This setting will toggle from Yes to No by pressing the Recirc
Option button. Once the option is set press the [CONFIRM] key and the X
will change from gray to blue. Select Yes or No as dictated by facility
management. Record this setting on Installation Checklist. If this option is
set to Yes, perform the following:
-
RECIRC GOAL: Set the recirculation goal by selecting the Recirc
Goal data button then use the [
volume from 10 to 9990ml. When the desired amount has been reached,
press [CONFIRM] key. Record this setting on the Installation Checklist.
RECIRC TIME: Set the recirculation time by selecting the Recirc Time
-
data button the use the [
min to 9 hrs 59 min. When the desired amount has been reached, press
[CONFIRM] key. Record this setting on the Installation Checklist.
=
>
] and [
] keys to select the desired volume from 0 to 9.9
=
>
] and [
] keys to select the desired volume from 0 to
=
>
] and [
] keys to select the desired volume from 0 to 1000ml.
=
=
>
] and [