The following procedure calibrates the pressure sensor on the K arterial blood
pump.
-
With the machine turned OFF, position the blood pump Service Jumper into
the In Service Mode position.
-
Locate DIP switch 2 on the blood pump LP955 board and configure it as
follows:
• For prepump, DIP switch 2 must be in the ON position
• For postpump, DIP switch 2 must be in the OFF position
-
With the 2008K machine in Service mode, turn the power ON.
-
When the power up sequence is complete, the blood pump display will be
alternating between these two displays:
N
Note: If the above display has a number 2 instead of a number 1, press the
>
[
] down key on the blood pump control panel to change it to 1.
Page 26
Fresenius 2008K Calibration Procedures
P/N 507296 Rev. A