HP 16550A User Reference page 93

100-mhz state/500-mhz timing logic analyzer
Hide thumbs Also See for 16550A:
Table of Contents

Advertisement

1
Example
See
Also
The Trigger Menu
Modifying the User-level Macro
Using the Timer
Timers are like other resource terms in that they are either true or false.
Timers can be set to Start, Stop, Pause, or Continue as the analyzer enters a
sequence level. The two timers are global, so each sequence level has the
ability to control the same timer. The default timer condition in
all
sequence
levels is Off.
The following statement assigns the
"Timer
1" to the secondary
"else
on"
branch.
If
the bit pattern is not seen before the timer counts 400ns, the
secondary branch is taken to level 2.
TRIGGER on ( _ _ _ _ _ _ _ _ a _ _ _ _ _ _ _ _
J (
0 cc u rs ) (.._ _____
J
E 1 s e on
(..._ _ _ _ _ _
r_i
m_e_. r_1_>_.:::i_o_c_,
n_1
s _ _ _ _ _
J
go to 1
eve
1 (
2
J
The "Timer Terms" found later
in
this chapter for information on how timers
work and how to assign a value to a timer.