HP 16550A User Reference page 237

100-mhz state/500-mhz timing logic analyzer
Hide thumbs Also See for 16550A:
Table of Contents

Advertisement

I
Error Messages
Warning Messages
Two pod pairs are needed to use both timers.
If
both timers are
being used in a single chip analyzer, and there are no spare chips, this
message is shown when Run is selected. Timer2 references in the
sequencer
will
not be valid.
No active analyzer. This message is displayed if
Run
is selected with
no analyzers turned on.
Cannot read unrecognized data. The user is trying to translate a SP A
configuration from a foreign module.
Demultiplexed clocking cannot be translated. The user is trying to
translate a configuration that has demultiplexed or mixed clocks.
User thresholds have been truncated. The user is trying to translate
a configuration having thresholds greater than 6 Volts or less than -6
Volts.
Slave clocks may need manual adjustment. The user is trying to
translate an HP 16540/41A,D configuration which has slave clocks
specified. This message appears when the translation is not direct.
Clock Qualifiers not fully restored. The user is trying to translate a
configuration that has more than two qualifiers.
No state machines for this module. The user is trying to load an
inverse assembler into a module with no state machines.
Error loading DISPLAY!. User is loading a configuration with a
corrupt DISPLAY! section.
Pods have been truncated. User is loading a configuration with
number of chips greater than the current system's number of chips.
Clock pod and least significant pods have been preserved. User is
loading a configuration with number of chips greater than the current
system's number of chips
Need to have two free sequence levels. For each sequence level
with a "<"assigned, you must leave two sequence levels free. To free up
two levels, simply delete two levels.
Mixed Mode not available User is trying te show mixed mode
incorrectly.
11-6

Advertisement

Table of Contents
loading

Table of Contents