HP 16550A User Reference page 197

100-mhz state/500-mhz timing logic analyzer
Hide thumbs Also See for 16550A:
Table of Contents

Advertisement

I
The Chart Menu
Stop Measurement Field
X-0
The X-0 option is available when the Count field in the Trigger menu is set to
Time.
When X-0 is selected, a repetitive run is stopped when a comparison of the
time period between the X and 0 markers and one of the following time
period options is true:
Less Than X-0 time must be less than the time value that you specify
in the Time field.
Greater Than X-0 time must be greater than the time value that you
specify in the Time field.
In Range X-0 pattern must be within the time range value that you
specify in the two Time fields.
Not in Range X-0 pattern must not be within the time range value that
you specify in the two Time fields.
--
~
~
O Marker>
( 1919 )
X-0 type field
Stop measurement (
X-0
)is • • • •
(
Clear
Pat tern
X-0 Field
8-20
X-0 type options
Not in range