HP 16550A User Reference page 235

100-mhz state/500-mhz timing logic analyzer
Hide thumbs Also See for 16550A:
Table of Contents

Advertisement

I
Error Messages
Error Messages
Cannot insert after last sequence level. Sequence levels cannot be
added after the last level. Add levels before the last level.
Last level of state sequence must be a user level. The last level of
a state sequence must be a user macro level.
If
the user tries to select a
new macro or replace this level, this error results.
Cannot delete last level in state sequence. The last level in a state
sequence cannot be deleted.
Too many sequence levels needed! Choose "Cancel" to leave the
menu. This message is displayed when changes in a sequence level menu
causes the translation too use more than the maximum number of levels
available.
Too many sequence levels needed! This message is displayed when
using HP-IB commands and when changes in a sequence level menu
causes the translation too use more than the maximum number of levels
available.
Error: Min time is
=
Max time. This error message is displayed if the
user attempts to exit the pattern width violation macro level menu with
the
min
width time
=
to the max width time.
Macro requires timer resource assigned to this machine.
If
the
user tries to select the "event too soon after event", or the "event too late
after event" macro, and there are no timer resources assigned to the
current machine, the selection is ignored and the user is given this error
message.
This macro cannot be selected when no labels are turned on.
A
macro has been selected that has a label entry, but has no labels turned
on in the Format menu. Make sure the configuration is consistant across
menus.
Warning! Turning off this label will default trigger sequence!
This message appears if a label is used in a macro, and then it is turned
OFF in the Format menu. Make sure the configuration is consistant
across menus.
This macro requires two unused patterns assigned to this machine.
If
the user tries to select then-bit serial pattern macro and there are not at
least two pattern resources assigned to the machine that are "unused"
(defined as BOTH
all
don't cares AND not used in the sequence) then this
error results, and the selection is ignored.