Dc Current Specifications - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

1-33. DC Current Specifications

Resolution
Range
nA
99 % Confidence Level
220 µA
0
1
2.2 mA
10
22 mA
µA
[1]
220 mA
0.1
[1]
2.2 A
1
5725A Amplifier:
11 A
10
95 % Confidence Level
nA
220 µA
0.1
1
2.2 mA
10
22 mA
µA
[1]
220 mA
0.1
[1]
2.2 A
1
5725A Amplifier:
11 A
10
Note:
Maximum output from the calibrator's terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA ranges are increased
by a factor of 1.3 when supplied through 5725A terminals.
Specifications are otherwise identical for all output locations.
1.
Add to uncertainty specifications:
2
±200 x I
ppm for >100 mA on 220 mA range
2
±10 x I
ppm for >1 A on 2.2 A range
For fields strengths >0.4 V/m but ≤3 V/m, add 1 % of range.
2.
5720A Series II DC Current Specifications
Absolute Uncertainty
±5 °C from calibration temperature
24 Hours
90 Days
180 Days
40 + 7
42 + 7
30 + 8
35 + 8
30 + 50
35 + 50
40 + 0.8
45 + 0.8
47 + 0.8
60 + 15
70 + 15
330 + 470
340 + 480
350 + 480
32 + 6
35 + 6
25 + 7
30 + 7
25 + 40
30 + 40
35 + 0.7
40 + 0.7
42 + 0.7
50 + 12
60 + 12
330 + 470
340 + 480
350 + 480
Introduction and Specifications
Relative Uncertainty ±1 °C
[2]
1 Year
± (ppm output + nA)
45 + 7
50 + 7
37 + 8
40 + 8
37 + 50
40 + 50
± (ppm output + µA)
50 + 0.8
80 + 15
90 + 15
360 + 480
± (ppm output + nA)
37 + 6
40 + 6
33 + 7
35 + 7
33 + 40
35 + 40
± (ppm output + µA)
45 + 0.7
70 + 12
80 + 12
360 + 480
Electrical Specifications
24 Hours
90 Days
24 + 2
26 + 2
24 + 5
26 + 5
24 + 50
26 + 50
26 + 0.3
30 + 0.3
40 + 7
45 + 7
100 + 130
110 + 130
20 + 1.6
22 + 1.6
20 + 4
22 + 4
20 + 40
22 + 40
22 + 0.25
25 + 0.25
32 + 6
40 + 6
100 + 130
110 + 130
1
1-33