Verifying The 22 Ma Range Of The 22 Ma/220 Ma Amp Circuit - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

the shunt and output switching is correct check Q4, Q5, K3 and their associated
components in the 220 µA /2.2 mA AMP circuit.
6. Check the 22 mA Range of the 22 mA/220 mA AMP. Set the Calibrator to 20 mA,
standby. Set the external ac reference to 6.3V (±0.1V) at 1 kHz. Using a DMM
measure the ac voltage at the collector of Q6 and verify it is 2.0V ac ±10%. If a
failure is detected, first verify the 100Ω shunt on HR2 is correct and is connected to
SCOM through relays in K8C, K7A, K5D, K10, K13, and 1.2Ω load resistor R14, as
shown in Figure 5-10. If the shunt and output switching are correct, check Q6-Q13,
K4, and the associated components in the 22 mA/220 mA AMP circuit.
2V ac
±10%
Figure 5-9. Verifying the 2.2 mA Range of the 220 µA /2.2 mA Amp Circuit
2V ac
±10%
Figure 5-10. Verifying the 22 mA Range of the 22 mA/220 mA Amp Circuit
7. Check the 220 mA Range of the 22 mA/220 mA AMP. Set the Calibrator to 200 mA,
standby. Set the external ac reference to 4.6V (±0.1V) at 1 kHz. Using a DMM
measure the ac voltage at the collector of Q6 and verify it is 2.0V ac ±10%. If a
failure first verify the 10Ω shunt on HR2 is correct and it's connected to SCOM
through relays in K9C, K8C, K7A, K5D, K10, K13, and 1.2Ω load resistor R14 as
shown in Figure 5-11. If the shunt and output switching are correct, check Q6-Q13,
K4, and the associated components in the 22 mA/220 mA AMP circuit.
8. Check the NEG. FB BUFFER. Set the Calibrator to 200 mA, standby. Set the
external ac reference to 4.6V at 1 kHz. Using a DMM measure the ac voltage at the
collector of Q6 and Note the reading. Next, measure the voltage at TP3 and verify it
is the same as the previous Noted reading ±0.01%. If a failure is detected, check U2,
K5A, K6A, K8A,K9A and the associated components.
10 k
HR2
SHUNT
K7C, K5D,
K10, K13
10 k
HR2
SHUNT
K8C, K7A,
K5, K10, K13
Troubleshooting
Component-level Troubleshooting
1.2
R14
SCOM
F5-9.EPS
1.2
R14
SCOM
F5-10.EPS
5
5-49