Fluke 5700A Series Service Manual page 274

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
Output
Current
2 mA
2 mA
2 mA
2 mA
2 mA
2 mA
200 µA
200 µA
200 µA
200 µA
200 µA
200 µA
1: The spec is modified because of the capacitive loading of the 2K resistor by the cable and input of the
5790A.
Output
Current
2 mA
2 mA
2 mA
2 mA
2 mA
2 mA
200 µA
200 µA
200 µA
200 µA
200 µA
200 µA
1: The spec is modified because of the capacitive loading of the 2K resistor by the cable and input of the
5790A.
3-64
Table 3-28. 5720A AC Current 2 mA and 200 µA Accuracy Test Record
Frequency
Error
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
Table 3-29. 5700A AC Current 2 mA and 200 µA Accuracy Test Record
Frequency
Error
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
5790A
Spec.
Transfer
± ppm
Spec
± ppm
(90-day)
305
210
200
73
150
27
150
27
285
27
1500
27
380
210
240
73
180
27
180
27
395
27
1500
27
5790A
Spec
Transfer
± ppm
Spec
± ppm
(90-day)
725
210
400
73
160
27
160
27
850
27
2100
27
850
210
505
73
240
27
240
27
850
27
2100
27
Test Limits
± ppm
TUR
GB
Factor
1.45
0.874
267
2.74
0.971
194
5.56
1.000
150
5.56
1.000
150
10.56
1.000
285
55.56
1.000
1500
1.81
0.916
348
3.29
0.988
237
6.67
1.000
180
6.67
1.000
180
14.63
1.000
+365/-425 (1)
55.56
1.000
+1380/-1620
(1)
Test Limits
± ppm
TUR
GB
Factor
3.45
0.991
718
1.000
400
5.48
5.93
1.000
160
5.93
1.000
160
31.48
1.000
850
77.8
1.000
2100
4.05
1.000
850
1.000
505
6.92
8.89
1.000
240
8.89
1.000
240
31.48
1.000
+820/-880 (1)
77.8
1.000
+1980/-2220
(1)

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series

Table of Contents