8. Reverse the HI and LO connections at the 732B terminals, and press E. The
following displays appear, asking you to wait before proceeding with the 10 kΩ
standard.
PREV
MENU
After a few seconds, the following display appears:
PREV
MENU
9. Connect the calibrator to the 10 kΩ standard as shown in Figure 3-4 and enter the
true value of the standard. If the standard is not between 9 kΩ and 11 kΩ, an error
message appears, which allows you to start over from this point with a different
standard. Press E again to open the following display:
PREV
MENU
When the internal 10 kΩ reference has been characterized, the following message
appears, which lets you accept or reject the changes that are about to be made to the
calibration constant:
PREV
MENU
Calibrating the 5700A/5720A Series II to External Standards
Calibration and Verification
IF3-8.EPS
IF3-9.EPS
IF3-10.EPS
IF3-11.EPS
3
3-11