Determine Specification Limits for other Calibration Intervals

The verification procedures in this document test to the 90 day 99 % confidence
specification limits. For other calibration intervals, it is necessary to calculate
limits based upon the specification that was selected. The subsequent examples
show how the 90 day specification limits were calculated. These examples
illustrate how to calculate the specifications limits for other intervals or levels of
confidence.
The next example shows how to calculate a specification limit for a particular test
point from the combined specification which is listed as a percentage (or parts
per million) of reading plus a floor error in microvolts, millivolts, microamps or
nanoamps. The component of the specification in parts per million is referred to
as a range specification. The floor specification has the same base units of
measurement as the output value (volts, amps). It is not appropriate to add a
range specification and floor specification together directly, because they are not
in the same units of measurement. In order to add the two quantities, it is
48
Table 3. List of Required Equipment for Main Output
Fluke 5790A
Fluke 742A Series
Fluke 752A
Fluke 8508A
Fluke 732B
Tektronix FCA3000
Fluke A40B Series
Guildline 9230 0.1 Ω and
0.01 Ω
Fluke 8508A
5730A-7003
Equipment Required for Wideband AC Module (Option 5700-03) Verification
Fluke 5790A with Wideband
Option -03
Fluke Cable
Fluke Termination
Pomona 1269 BNC(F) to dual
banana plug
Kings KN-99-46 N(F) to
BNC(M)
Description
Application
ACV, ACI
Cal, Ohms, DCI
DCV
DCV
DCV
Frequency
ACI
DCI
Ohms, DCI, ACV, and DCV
Various
Wideband
Wideband
Wideband
Wideband
Wideband