General Specifications - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

1-28. General Specifications

Warm-Up Time....................................................... Twice the time since last warmed up, to a maximum of 30 minutes.
System Installation ............................................... Rear output configuration and rack- mount kit available.
Standard Interfaces .............................................. IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in
Temperature Performance
Operating ............................................................ 0 °C to 50 °C
Calibration........................................................... 15 °C to 35 °C
Storage ............................................................... -40 °C to 75 °C
Relative Humidity
Operating ............................................................ <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C
Storage ............................................................... <95 %, non-condensing. A power stabilization period of four days may
Safety ..................................................................... Designed to comply with UL3111; EN61010; CSA C22.2 No. 1010;
Guard Isolation...................................................... 20 V
EMI/RFI................................................................... Designed to comply with FCC Rules Part 15, Subpart B, Class B;
ElectroStatic Discharge........................................ This instrument meets criteria C for ESD requirements per EN61326
Line Power
Line Frequency .................................................. 47 to 63 Hz; ±10 % 100 V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V,
Maximum Power
5700A/5720A .................................................. 300 VA
5725A ............................................................. 750 VA
Weight
5700A/5720A ...................................................... 27kg (62 lbs)
5725A.................................................................. 32kg (70 lbs)
Size
5700A/5720A
Height ............................................................. 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet
Width............................................................... 43.2 cm (17 in), standard rack width
Depth .............................................................. 63.0 cm (24.8 in), overall; 57.8 cm (22.7 in), rack depth
5725A
Height ............................................................. 13.3 cm, (5.25 in)
Width and Depth ............................................. Same as 5700A/5720A. Both units project 5.1 cm, (2 in) from rack
(BNC), phase reference out (BNC).
be required after extended storage at high temperature and humidity.
ANSI/ISA S82.01-1994
EN50081-1, EN50082-1
240 V
front.
Introduction and Specifications

General Specifications

1
1-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series

Table of Contents