Fluke 5700A Series Service Manual page 220

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
6. Enter the true value of the 732B 10V output. The true value is the value printed on
7. To reject the changes, return to the calibration menu shown in step 2 by pressing
3-10
the calibration sticker, plus or minus, as well as drift that has occurred since
calibration. That drift can be estimated provided control charts have been maintained
for the 732B.
If the entered value is not between 9V and 11V, and error message appears, which
lets you start over from this point with a calibrated 732B. Press E, and the
display shows the following:
PREV
MENU
When the calibrator's 6.5V and 13V references have been characterized, the display
shows the following message, which lets you accept or reject the changes that are
about to be made to the calibration constants.
PREV
MENU
P. Otherwise, press the softkey under "Proceed" to accept and save the changes,
and to open the display shown below, letting you continue with calibration.
PREV
MENU
IF3-5.EPS
IF3-6.EPS
IF3-7.EPS

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series

Table of Contents