Fluke 5700A Series Service Manual page 257

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

Item
Description
No.
1.
Voltage Reference
2.
Resistance Standards
3.
Reference Voltage
Divider
4.
Null Detector
5.
Low Thermal EMF
Cables
6.
Digital Multimeter
7.
Current Source
8.
AC Measurement
Standard
Table 3-15. Minimum Use Requirements
Minimum Use Specifications
Calibration Equipment
10V nominal, true value certified to within
±1.5 ppm
1Ω nominal, true value certified to within 10
ppm, 10 kΩ nominal, true value certified to
within 4 ppm
Calibration Verification Equipment
Range uncertainty 100:1, 1 kV input, ±0.5
ppm 10:1 100V input ±0.2 ppm
Leakage resistance to case: 10
Resolution: 3 µV full scale
Plug-in. Copper or gold-flashed copper (two
cables per set, two sets required).
Spade lug. Copper or gold-flashed copper
(two cables per set, two sets required).
DC Voltage Range: 0.1 to 10V Resolution
and short-term stability: ±2 ppm Resistance
range: 1Ω to 10 MΩ Resolution and short-
term stability :±20 uΩ at 1Ω, 1.9Ω ±5 ppm at
10Ω, 19Ω ±2 ppm at 100Ω to 1.9 MΩ ±4
ppm at 10 MΩ, 19 MΩ
Range: 10 mA and 100 mA Typical short-
term stability ±15 ppm for 5 minutes
Ranges: 2.2 mV through 1000V AC
Frequency: 10 Hz to 1 MHz
24 to 7500 ppm, depending on amplitude
and frequency (see Table 3-9.)
Wideband Ranges: 2.2 mV through 7V
Frequency: 10 Hz to 30 MHz
Uncertainty: 0.03% to 0.9%, depending on
amplitude and frequency (see Table 3-18.)
Calibration and Verification
Minimum Use Requirements
Recommended
Equipment
Fluke 732B
Fluke 742A Series, 1Ω
and 10 kΩ
Fluke 752A
EM Electronics N11
Ω min.
12
(Fluke 845A)
Fluke 5440A-7002
Fluke 5440A-7003
Wavetek 1281 or
HP 3458A
Fluke 55X0A,
5700A/5720A Series II,
Fluke 5790A (Option -03
required for Wideband
Flatness Verification)
3
3-47

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series

Table of Contents