1-35. Wideband AC Voltage (Option 5700-03) Specifications

Specifications apply to the end of the cable and 50 Ω termination used for calibration.
Range
Volts
dBm
1.1 mV
-46
3 mV
-37
11 mV
-26
33 mV
-17
110 mV
-6.2
330 mV
+3.4
1.1 V
+14
3.5 V
+24
Frequency
Frequency
Resolution
(Hz)
(Hz)
10 - 30
0.01
30 - 120
0.01
120 - 1.2 k
0.1
1.2 k - 12 k
1
12 k - 120 k
10
120 k - 1.2 M
100
1.2 M - 2 M
100 k
2 M - 10 M
100 k
10 M - 20 M
1 M
20 M - 30 M
1 M
Additional Operating Information:
dBm reference = 50 Ω
Range boundaries are at voltage points, dBm levels are approximate.
Power
) ; 0.22361 V across 50 Ω = 1 mW or 0 dBm
dBm = 10 log (
1
mW
Minimum Output ................................................... 300 µV (-57 dBm)
Frequency Uncertainty ......................................... ±0.01 %
Frequency Resolution .......................................... 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz
Overload Protection.............................................. A short circuit on the wideband output will not result in damage. After
Resolution
24 Hours
10 nV
0.4 + 0.4
10 nV
0.4 + 1
100 nV
0.2 + 4
100 nV
0.2 + 10
1 µV
0.2 + 40
1 µV
0.2 + 100
10 µV
0.2 + 400
10 µV
0.15 + 500
Amplitude Flatness, 1 kHz Reference
Voltage Range
1.1 mV
3 mV
± (% output + floor indicated)
0.3
0.3
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.2 + 3 µV
0.1 + 3 µV
0.2 + 3 µV
0.1 + 3 µV
0.4 + 3 µV
0.3 + 3 µV
0.6 + 3 µV
0.5 + 3 µV
1.5 + 15 µV
1.5 + 3 µV
settling time, normal operation is restored upon removal.
Introduction and Specifications
Absolute Uncertainty
±5 °C from calibration temperature
30 Hz - 500 kHz
90 Days
± (% output +µV)
0.5 + 0.4
0.45 + 1
0.35 + 4
0.3 + 10
0.3 + 40
0.25 + 100
0.25 + 400
0.2 + 500
Temperature
Coefficient
>3 mV
±ppm/°C
0.3
100
0.1
100
0.1
100
0.1
100
0.1
100
0.1 + 3 µV
100
0.1 + 3 µV
100
0.2 + 3 µV
100
0.4 + 3 µV
150
1 + 3 µV
300
Electrical Specifications
180 Days
1 Year
0.6 + 0.4
0.8 + 2
0.5 + 1
0.7 + 3
0.5 + 4
0.7 + 8
0.45 + 10
0.6 + 16
0.45 + 40
0.6 + 40
0.35 + 100
0.5 + 100
0.35 + 400
0.5 + 400
0.3 + 500
0.4 + 500
Settling Time
Harmonic
To Full
Distortion
Accuracy
(dB)
(Seconds)
7
-40
7
-40
5
-40
5
-40
5
-40
5
-40
0.5
-40
0.5
-40
0.5
-34
0.5
-34
1
1-41