Output
Current
2A
2A
2A
2A
200 mA
200 mA
200 mA
200 mA
200 mA
200 mA
20 mA
20 mA
20 mA
20 mA
20 mA
20 mA
10A (1)
10A (1)
10A (1)
10A (1)
1: If a 5725A is attached.
3-62
Table 3-26. 5720A AC Current 20 mA to 10A Accuracy Test Record
2A, 200 mA, 20 mA and 10A Test Record
A40 Shunt
Freq
AC-DC Diff
40 Hz
1 kHz
5 kHz
10 kHz
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
10 Hz
20 Hz
40 Hz
1 kHz
5 kHz
10 kHz
40 Hz
1 kHz
5 kHz
10 kHz
UUT Error
Calculated
Display
Error
Spec
± ppm
320
320
510
7100
305
200
145
145
240
1160
305
200
150
150
255
1400
417
417
888
3375