Rear Panel Assembly Access - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

4-10. Rear Panel Assembly Access

Refer to Figure 4-3 during the following procedure:
1
REMOVE FACK SCREWS TO SEPARATE THE REAR PANEL CIRCUIT BOARD FROM
THE METAL HOUSING.
1. Remove the six securing screws for the Rear Panel assembly housing.
2. Gently pull the rear panel housing from the Rear Panel.
3. Allow the rear panel housing to lay flat on the work surface by removing the two
ribbon cables from the Rear Panel board.
4. Remove the two nuts at TB1 and TB2 on the paddle board; separate the associated
wires from the paddle board.
Figure 4-2. Rear Panel Removal
REMOVE
1
1
REMOVE
Figure 4-3. Rear Panel Assembly Access
REMOVE
1
1
Maintenance
4
Access Procedures
F4-3.EPS