3.
Rosemount 3051CD0 is only available with output code A and X. For output code A, only process flange code 0 (alternate flange H2, H7, HJ or HK),
isolating diaphragm code 2, O-ring code A and bolting option L4 are available. For output code X, only process flange code 0 (alternate flange H2),
isolating diaphragm code 2, O-ring code A and bolting option L4 are available.
4.
Option HR5 configures the HART output to HART Revision 5. Option HR7 configures the HART output to HART Revision 7. The device can be field
configured to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
5.
For local addressing and configuration, M4 (LOI) is required.
6.
Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety (option code I1), IECEx
Intrinsic Safety (option code I7) and EAC Intrinsic Safety (option code IM).
7.
Only available with C6, E2, E5, I5, K5, KB and E8 product certifications. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, CT.
8.
Materials of Construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental
limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
9.
Not available with Wireless Output (code X).
10. Only available with Wireless Output (code X).
11. Not available with Product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, KD.
12. Transmitter conduit entry will be
13. Only valid with F
Fieldbus output code F.
OUNDATION
14. Only available with HART 4-20 mA output (code A).
15. Requires 0 code in materials of construction for alternate process connection.
16. Not valid with option code P9 for 4500 psi static pressure.
17. "Assemble-to" items are specified separately and require a completed model number.
18. Process flange limited to coplanar (option codes 2, 3, 5, 7, 8) or traditional (option codes H2, H3, H7).
19. Not valid with option code D9 for RC
20. Not valid for option codes DF and D9 for Adapters.
21. Panel mounting bolts are not supplied.
22. Dust approval not applicable to output code X. See
23. Only available with output codes A - 4-20mA HART, F - F
24. Nonincendive certification not provided with Wireless output option code (X).
25. Not available with Alloy C-276 isolator (code 3), tantalum isolator (code 5), all cast C-276 flanges, all plated CS flanges, all DIN flanges, all Level flanges,
assemble-to manifolds (codes S5 and S6), assemble-to seals (codes S1 and S2), assemble-to primary elements (codes S3 and S4), surface finish
certification (code Q16), and remote seal system report (code QZ).
26. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, N7
27. Not available with F
OUNDATION
28. Contact an Emerson Process Management representative for availability.
29. Only available with HART 4-20 mA Output (output code A) and Wireless Output (output code X)
30. The T1 option is not needed with FISCO product certifications; transient protection is included in the FISCO product certification codes IA, IB, and IE.
31. Not valid with alternate process connection options S3, S4, S5, and S6.
32. Transmitter is shipped with a 316 SST conduit plug (uninstalled) in place of standard carbon steel conduit plug.
33. Not available with alternate process connection; DIN flanges and level flanges.
34. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
35. NACE compliant wetted materials are identified by
Specifications and Reference Data
1
1
1
/
2
NPT and a
/
2
NPT to G
/
2
thread adapter will be provided.
1
/
Adapters.
2
"Rosemount 3051 Wireless Pressure Transmitter" on page 146
OUNDATION
Fieldbus (output code F), Wireless (output code X), or Low Power output (output code M).
Footnote
8.
Specifications and Reference Data
Fieldbus, and W - PROFIBUS PA. Also only available with G
August 2016
for Wireless approvals.
1
/