Types Of Signal Sources; Floating Signal Sources; Ground-Referenced Signal Sources; Differential Measurements - National Instruments PCI-6110E/6111E User Manual

Multifunction i/o boards for pci bus computers
Table of Contents

Advertisement

Types of Signal Sources

Floating Signal Sources

Ground-Referenced Signal Sources

Differential Measurements

© National Instruments Corporation
When making signal connections, you must first determine whether the
signal sources are floating or ground-referenced. The following
sections describe these two types of signals.
A floating signal source is not connected in any way to the building
ground system but, rather, has an isolated ground-reference point. Some
examples of floating signal sources are outputs of transformers,
thermocouples, battery-powered devices, optical isolator outputs, and
isolation amplifiers. An instrument or device that has an isolated output
is a floating signal source. You must tie the ground reference of a
floating signal to the 611X E board analog input ground to establish a
local or onboard reference for the signal. Otherwise, the measured input
signal varies as the source floats out of the common-mode input range.
A ground-referenced signal source is connected in some way to the
building system ground and is, therefore, already connected to a
common ground point with respect to the 611X E board, assuming that
the computer is plugged into the same power system. Nonisolated
outputs of instruments and devices that plug into the building power
system fall into this category.
The difference in ground potential between two instruments connected
to the same building power system is typically between 1 and 100 mV
but can be much higher if power distribution circuits are not properly
connected. If a grounded signal source is improperly measured, this
difference may appear as an error in the measurement. The connection
instructions for grounded signal sources are designed to eliminate this
ground potential difference from the measured signal.
The following sections discuss the use of differential (DIFF)
measurements and considerations for measuring both floating and
ground-referenced signal sources.
4-9
Chapter 4
Signal Connections
PCI-6110E/6111E User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Daq pci-6110eDaq pci-6111e

Table of Contents