Analog Input Signal Connections - National Instruments PCI-6110E/6111E User Manual

Multifunction i/o boards for pci bus computers
Table of Contents

Advertisement

Chapter 4
Signal Connections

Analog Input Signal Connections

Caution:
!
PCI-6110E/6111E User Manual
The analog input signals for the 611X E board are ACH<0..3>+ and
ACH<0..3>–. The ACH<0..3>+ signals are routed to the positive input
of the PGIA, and signals connected to ACH<0..3>– are routed to the
negative input of the PGIA.
Exceeding the differential and common-mode input ranges distorts your
input signals. Exceeding the maximum input voltage rating can damage
the 611X E board and the computer. National Instruments is
any damages resulting from such signal connections. The maximum input
voltage ratings are listed in the Protection column of Table 4-2.
With the different configurations, you can use the PGIA in different
ways. Figure 4-2 shows a diagram of your 611X E board PGIA.
V
in+
V
in-
The PGIA applies gain and common-mode voltage rejection and
presents high input impedance to the analog input signals connected to
the 611X E board. Signals are routed to the positive and negative inputs
of the PGIA. The PGIA converts two input signals to a signal that is the
difference between the two input signals multiplied by the gain setting
of the amplifier. The amplifier output voltage is referenced to the
ground for the board. The 611X E board A/D converter (ADC) measures
this output voltage when it performs A/D conversions.
Instrumentation
+
PGIA
-
V
= [V
- V
m
in+
in-
611
E
Figure 4-2.
X
Board PGIA
4-8
liable for
NOT
Amplifier
+
V
Measured
m
Voltage
-
]* Gain
© National Instruments Corporation

Advertisement

Table of Contents
loading

This manual is also suitable for:

Daq pci-6110eDaq pci-6111e

Table of Contents