Function Check With Control Solution - Beurer BGL60 Instructions For Use Manual

Table of Contents

Advertisement

6.4

Function check with control solution

The control solution is used for checking the whole blood sugar measuring sys-
tem. This enables you to determine whether the measuring device and the test
strips are functioning correctly together and whether the test is performed cor-
rectly. It is very important that you perform these tests using the control solution
to ensure accurate measuring results.
Important
Never use a control solution from other manufacturers. You can only verify the
proper function of your measuring device using the Beurer control solution.
When is it recommended to perform a test using the control solution?
• After opening a new test strip container.
• If you suspect that the measuring device or the test strips could be defective.
• If your blood sugar measurement values do not correspond to the way you are
feeling.
• If you have dropped the measuring device or it has been subjected to other
mechanical strain.
• Every time you suspect that the blood sugar results may be incorrect.
It is sufficient to perform a single control solution test if the result is within the
recommended range.
Warning
Do not ingest the control solution. The control solution is only to be used for
function tests and is for external use only.
38