4.2.3.8 AF Motor Assembly Removal

This section describes how to remove the motor which drives AF drive
mechanism.
1. Remove the glass frame assembly. (See Section 4.2.3.5.)
2. Remove 4 screws (No.3) and remove the cover which covers the AF
drive mechanism.
3. Remove 2 screws (No.6) and remove the tension plate assembly,
spring plate and fixing tension plate to loosen the timing belt.
4. Disconnect the cable for the motor from the relay connector, then
remove 4 screws (No.4) securing the motor and remove the motor.
Tension Plate Assembly
Figure 4-22. Tension Plate Assembly Removal
Rev. A
Screws (No.6)
Fixing Tension Plate
Spring Plate
Screws (No.4)
Figure 4-23. AF Motor Assembly Removal
CAUTION
You are required to set the timing belt with a
special care while installing the AF motor
assembly. Be sure to follow the instruction
provided in the next page to set the timing belt
peoperly.
Disassembly and Assembly
Motor Assembly (AF)
4-17