Invacare
- STORM
8.2.5
Replacing the lighting PCB
Replacement is described in the following section based on the example of a mobility aid with an
optional lighting PCB and an optional actuator module.
Requirements:
· 4 mm Allen key
· 8 mm socket spanner
Dismantling the lighting PCB:
·
If possible, use the lifter to move the seat to the uppermost position.
·
Remove the rear and the centre cover as described in chapters 8.1.1 and 8.1.2.
·
Carefully note the location of the cable and the connection locations of the various plugs. Either
mark each plug and socket, or take a photograph with a digital camera.
·
The pictures show the power module (4), the optional lighting PCB (2) and the optional actuator
modules (1) & (3).
·
Pull the plug (1) out of the lighting PCB.
·
Undo the two Allen screws (2) with a 4 mm Allen key.
·
Lift off the lighting PCB together with the retaining bracket.
4
SERVICE MANUAL
49