8.16.2 Replacing the EasyAdapt back

Requirements:
· 5 mm Allen key
· Oblique pliers
· Cable ties
Dismantling the back unit:
·
Remove the back cushion.
·
Pull the actuator cable out of the actuator module under the seat.
·
Pull the cable out of the mobility aid. Remove any cable ties which are fitted.
·
Loosen and remove the Allen screws (5) incl. the washers on the left and right side of the seat.
·
Carefully remove the back section (1) together with the lower part (2) from the seat support
(3). The two parts are interconnected via the arm (4).
Refitting the back unit:
·
Refit the parts in reverse order.
·
To conclude, check all the mobility aid functions.