10 – Camera program
Command
Settings
Point to Settings and press the joystick to display a Line settings dialog box where
you can change the settings for the line.
10588803;a2
Figure 10.27 Line dialog box
Figure 10.28 Explanations of the Line dialog box
Label
Local
Emissivity
10
Emissivity table
84
Explanation
See below.
Value
On
Off
User-defined
(0.01–1.00)
User-defined
Comments
Select On to set the emissivity, the reflected tem-
perature, and the distance for this line only.
Selecting On will also assign an asterisk to the
measurement marker's label.
You can set the Emissivity if Local is enabled. If
not, this option will be shaded.
➲ If you enter an emissivity value less than 0.30
the emissivity box will begin flashing to remind you
that this value is unusually low.
Press Emissivity table to display an emissivity
table on the screen.
You can use this emissivity table to find emissivities
for a number of different materials. An emissivity
table can be created and edited in FLIR Systems's
PC software.
➲ The emissivity file can be stored at root level or
at directory level. However, the camera software
prioritizes files that are stored at directory level and
the directory has to be selected in order to store
the emissivity file in the camera memory. If the
camera software does not find an emissivity file at
directory level, it searches for similar files at root
level and saves those instead.
Publ. No. 1557966 Rev. a155 – ENGLISH (EN) – February 6, 2006