Specifications - Nikon Eclipse LV100POL Instructions Manual

Polarizing microscope
Table of Contents

Advertisement

Specifications

Model
Optical system
Illumination system
Focusing mechanism
Eyepiece
Stage
Nosepiece
Input ratings
Power cord
Operating conditions
Storage conditions
62
ECLIPSE LV100POL
CFI60 system (chromatic aberration free infinity system)
Epi/dia selector type
Episcopic illumination:
Built-in type lamp power supply
Specified illuminator:
LV-UEPI Universal Epi Illuminator (NCB11, ND4, and ND16
are installed and can be replaced by users)
Diascopic illumination:
Built-in type lamp power supply, fly's eye lens, NCB11 and
ND8 are installed but cannot be replaced by users
Lamp ratings:
12 VDC, 50 W halogen lamp
Specified lamp:
LV-HL50W, 12 V, 50 W longlife halogen lamp
Specified lamp house:
LV-LH50PC precentered lamp house
Manual operation type single axis coarse/fine focus knob mechanism (calibration marking
for fine focus: 1 µm/marking)
Stroke:
30 mm with coarse focus stopper mechanism
Coarse focus knob:
14 mm/revolution
Fine focus knob:
0.1 mm/revolution
10x, field number: 22
Circular graduated stage
Provided with two verniers
Provided with a rotation mechanism with a clamp
Provided with 45° click-stop mechanism
Provided with two specimen holders
Quintuple type (five sockets)
Provided with objective centering mechanism
100-240 VAC ±10% 50/60 Hz
Input voltage:
Rated current:
1.2 A maximum
When the supply voltage is 100 V to 120 V
UL Listed detachable cord set, 3 conductor grounding Type
(3 conductor grounding Type SVT, No.18 AWG, 3m long
maximum, rated at 125V AC minimum)
When the supply voltage is 220 V to 240 V
Approved according to EU/EN standards, 3 conductor
grounding Type (3 conductor grounding Type H05VV-F, 3m
long maximum, rated at 250V AC minimum)
Temperature:
0°C to +40°C
Humidity:
85% relative humidity maximum (no condensation)
Altitude:
2000 m maximum
Pollution degree:
Degree 2
Installation category:
Category II
Electric shock protection class: Class 1
Indoor use only
Temperature:
-20°C to +60°C
Humidity:
90% relative humidity maximum (no condensation)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents