Rotate the nosepiece to locate the desired
17
magnification objective into the optical
path. Then, perform the microscopy.
• Each time you change objectives, the field
diaphragm and the condenser aperture
diaphragm on the epi illuminator must be
adjusted.
- The field aperture should be adjusted so that
its image circumscribes the view field.
- And the aperture diaphragm should be 70% to
80% of the numerical aperture of the
objective.
See Pages 43 and 44.
• Focus on the specimen again using the fine focus
knob or the coarse focus knob.
See Page 25.
• Use the ND filters on the epi illuminator to
adjust the brightness.
See Page 44.
• To use an oil immersion type objective or a
water immersion type objective, perform the oil
immersion or water immersion.
See Pages 34 and 35.
To perform the episcopic polarization
18
microscopy, locate the polarizer on the
side of the epi illuminator into the optical
path and locate the analyzer on the
polarizing intermediate tube into the
optical path.
• The polarizer azimuth should be adjusted with
the ring on the polarizer slider.
See Page 36.
To return to the bright-field microscopy
19
• Fully rotate the brightness control knob to the far side to turn off the epi illumination lamp.
• Pull out the illumination selector lever on the epi illuminator to set the dark-field (DF) position for it.
• Push the epi/dia selector switch to select the diascopic illumination. The indicator for the diascopic
illumination (lower side) turns on.
• Rotate the brightness control knob to turn on the diascopic illumination lamp and adjust the
brightness.
• Remove the analyzer out of the optical path.
Turn off all the power after completing the observation.
20
20
17
0
B F
D F
Switch to
17
any
desired
objective.
Ach r
0. 8
0.90
0.6
JAPAN
0.4
0.2
F.S.
17
18
0
B F
D F
Locate the polarizer
18
into the optical path.
0. 8
Ach r
0.90
0.6
JAPAN
0.4
0.2
F.S.
Adjust the field
diaphragm and the
aperture diaphragm.
A. ST O
P
F. ST O
P
NC B
ND 8
Readjust the focus.
Locate the analyzer
into the optical path.
P
A. ST O
P
F. ST O
NC B
ND 8