Operation of Each Part

1

Power ON / OFF

The power switch for the microscope is located beside
the AC inlet on the rear side.
To turn on the microscope, push the power switch to
the "|" side. To turn off the microscope, push the
power switch to the " " side.
When the power to the microscope is on, the power
indicator on the front side of the microscope is lit.
(When the lamp is off, the indicator is orange, and
when the lamp is on, the indicator is green.)
Power supply for the lamp
The LV100POL has a built-in power supply circuit for the halogen lamp. With the specified lamp
house (LV-LH50PC), the lamp lights up simultaneously as the power supply for the microscope is
turned on.
Power supply for the lamp
To use a lamp other than the specified one, turn on the power supply for the light source by
hand.
22
CAUTION !
- High Temperature -
1.
Do not touch the lamphouse while the lamp is lit.
The surface of the lamphouse becomes hot when
the lamp is on.
2.
Turn off the power and allow the lamp and lamp-
house to cool enough before replacing the lamp.
Wait for at least 30 minutes after turning off
the lamp
3.
Use 12V50W HALOGEN lamp only.
HALOGEN 12V50W
LAMP
0
LV-LH50PC
DC12V 50W
JAPAN
6 5 2 7 0 2
AC inlet
Power switch
ON
CLICK
JAPA
N
0.90
0. 2
Achr
0.4
0.6
Power
0.8
indicator
EPI /DIA
OFF