Specifications And Standards - Nikon ECLIPSE E600 POL Instructions Manual

Polarizing microscope
Table of Contents

Advertisement

46
Specifications and Standards
- •
1. Model Name:
2. Focusing Mechanism:
3. Stage:
4. Revolving Nosepiece:
5. Illumination:
~. •
6. Input Power Supply
'
Voltage:
7. Protection Class:
8. Operating Environmental
Conditions:
ECLIPSE E600 POL (microscope main body)
• Fine focus knob scale:
1 step equals I µm
• Amount of movement of fine focus knob:
One turn moves the stage up or down by 0.1 mm
• Amount of movement of coarse focus knob:
One turn moves the stage up or down by about 12 mm
• Range of stage vertical motion:
2 mm up and
23
mm down from the refrrence (focused)
position
• Circular graduated stage with two verniers
• Provided with a rotation mechanism with a clamp
• Provided with a 45° click-stop
• Provided with two specimen clips
•Quintuple
• Provided with an objective centering mechanism
• Internal Koehler-type diascopic illumination optics
• Provided with a PHOTO switch (5-level fine adjustment
selector)
• Lamp rating: I 2V DC. I OOW halogen lamp
(OSRAM HLX64623 or PHILIPS 7724)
• AC
I 00-
l 20V
±
I
Olk 50/60Hz
Current consumption:
Internal fuse rating:
• AC 230V
±
10% 50/60Hz
Current consumption:
Internal fuse rating:
Class I
o
to +40°C
2.4A or less
125V, 5A
l.2A or less
250V. T3.15A
• Temperature:
•Humidity:
• Altitude:
85% RH max., non-condensing
2000m max.
• Pollution:
Degree 2
• Installation:
Category II
• Indoor use only