Getting Started
8
Installation Instructions
Leveling the Weighing Platform Using the Level Indicator
At the place of installation, level the weighing platform using the leveling feet
so that the air bubble is centered within the circle of the level indicator. Check
to ensure that all 4 leveling feet securely touch the surface and that they equally
support the weight of the platform.
- Remove the load plate
- Loosen the locknuts on the leveling feet using a 17 mm open-end wrench
(spanner)
- Extend or retract the leveling feet using a 5 mm Allen wrench (key)
- After leveling the weighing platform, refasten the locknuts securely against
the platform frame
-
Replace the load plate