Control Solution - Infopia Finetest Auto-coding User Manual

Blood glucose monitoring system
Table of Contents

Advertisement

Automatic code recognition
Caution
1. Do not bend the test strip to prevent the Automatic code recognition failure.
2. If the code recognition label is damaged, the code recognition may fail. Please
check the code number on the LCD display window with the code number on
the test strip vial.
1 6
1. finetest Auto-coding
has automatic
TM
code recognition function.
2. It recognizes automatically the code
number. This is a very convenient
function preventing the inconvenience
of setting the code number on the
meter every time.
the
Checking the system with the finetest Auto-coding
The finetest Auto-coding
Control Solution is used to check that the meter and the
TM
test strips are properly working together to provide you with the most accurate
reading.
Note : The
Control Solution is sold separately. Low, normal and
finetest Auto-coding
TM
high level control solutions can be obtained through your local representative.
Control Solution Test can only be used with the
should be used during the following cases:
•When a new vial of test strips is opened.
•Any suspicion that the meter or test strips are not working properly.
•When your blood glucose test results are not consistent with your symptoms, or if you
think they are not accurate.
•If you drop the meter.
•Use routinely to obtain accurate results.
When the control solution is applied to the top edge of the
Strip, you should get results within the expected range printed on the label of the test
strip vial. If the control solution test results fall outside the range, repeat the test. Results
that fall out-side the range may be caused by:
•Error in performing the test.
•Failure to shake the control solution vial well enough (must shake vigorously).
•Failure to discard the first drop of control solution.
•Expired or contaminated control solution.
•Test Strip deterioration.
•Meter malfunction.
•Control solution that is too warm or cool.
Checking the system with

control solution

control solution
TM
System and
finetest Auto-coding
TM
finetest Auto-coding
Test
TM
1 7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Finetest Auto-coding and is the answer not in the manual?

Table of Contents

Save PDF