Performing a Glucose Control Solution Test

®
Element
NEO control solution contains a known amount of glucose that reacts with
®
Element
NEO test strip. By comparing your control solution test results with the expected
range printed on the test strip vial, you are able to check that the meter and test strips are
working properly, and that you are performing the test correctly. It is very important that you
perform routine checking to ensure accurate results.
The Glucose control solution should be used
- Before using a new box of test strips.
- Whenever you suspect the blood glucose test meter or blood glucose test strip is not
functioning properly.
- If your blood glucose test results are not consistent with your symptoms, or if you think
results are not accurate.
- If you have dropped the blood glucose test meter.
- For quality control in the point of care usage.
- For teaching or learning the system.
- When using the meter for blood glucose tests after any disinfection procedure.
Firmly insert the test strip into the meter test port. Insert
deeply in direction of arrow on test strip (arrow up). Please
do not insert the glucose test strip upside down.
When you insert the glucose test strip into the glucose test
meter, the power automatically adjusts the meter accordingly.
Press and release either the ▲ or ▼ button, then, Control
Solution bottle icon will appear. Then, press
(
).
• 16 •
1.Check the expiration date before performing a control
solution test. Do not use if expired. Please note the
expiration date marked on the control solution bottle.
2.Gently shake the control solution before use.
3.Discard the 1st drop of control solution. This will eliminate
any residue.
Place a drop of control solution liquid on a clean, dry surface
(e.g. the lid of the test strip vial)
4.Dip the test strip into the control solution liquid.
Results will appear in 5 seconds.
Compare the result to the range printed on the test strip
vial. Result should fall within that range.
CAUTION
1. Control solutions is for in-vitro diagnostic use only. This means that it is only used for
testing outside the body.
2. Control solution is exclusively made for the Element
System. Do not use other brand control solutions made by other manufacturers.
3. Check the expiration date on the vial. Do not use, if expired.
4. Once opened, use it for the next three months only. Record the discard date on the
control solution bottle. Discard after three months.
5. Do not swallow. This is NOT for human or animal consumption.
6. The control solution bottle icon contains letter 'C,' so as not to be confused with the
medication bottle. This will allow you to understand differences between a control
solution test and an actual blood test.
®
NEO Blood Glucose Monitoring