Checking the system with control solution

Quality control by control solution
Control testing, also known as quality control testing, is used to ensure all the
parts of the test system are working properly together and the test results are
accurate and reliable within the limits of the system. Contact customer support of
local representative.
The users should follow their facility's policies on when controls should be tested
(for example: with each new Test strip lot; monthly as a continued check on
storage conditions;whenever issues (storage,operator,or other) are identified or
there are questions about the results).
The control solution should be used
• Whenever you suspect the meter or test strip is not functioning properly.
• If your blood glucose test results are not consistent with your symptoms or if
you think they are not accurate.
• If you have dropped the meter
• For quality control at the point of care
• For teaching or learning the system
Note:
The LipidPro
TM
Control Solutions are sold separately.
The TC(Level 1, Level 2), TG(Level 1, Level 2) and HDL (Level 1, Level 2) control
solutions can be obtained.
• 22 •
Checking the system with control solution
Caution
Please check the expiry date on the instructions for use provided with each
control solution.
Total Cholesterol strip or Lipid profile test strip in Total Cholesterol test area
should be used only Total Cholesterol control level 1, level 2. HDL Choles-
terol strip or Lipid profile test strip in HDL Cholesterol test area should be
used only HDL Cholesterol control level 1, level 2.
Triglyceride strip or Lipid profile test strip in triglyceride test area should be
used only triglyceride control level 1, level 2.
Glucose test strip should be used only glucose control solution Low, Normal
and High.
It is recommended that the control solution is stored at room temperature
before testing.
• Lipid Profile test
Testing procedure
1. Press RFID(▼) button for 3 or more seconds.
2. Place the RFID-tag on the strip vial to the
area of the RFID symbol (
) on the meter.
Caution
The radiation of RFID (
) recognition is within 2cm.
The meter will revert back to stand-by status if there is no operation within 5 seconds.