Performing A Control Solution Test - Infopia Element compact User Manual

Blood glucose monitoring system
Table of Contents

Advertisement

Performing a Control Solution Test

1
1. Insert the test strip with the arrow facing up and pointing towards the test strip
port. The code number will appear on the screen,verify that the code number
displayed matches with the code number on the vial of test strips. If the two
code numbers do not match, call Customer Service at 1-888-446-3246.
2. Shake well before using. Discard the first drop and wipe the dispenser tip.
Squeeze a drop of control solution onto a clean, dry, non-absorbent surface.
Do not apply control solution to the test strip directly from the bottle. Replace
the lid on the control solution vial immediately after use.
41
2
3
4
3. Gently touch the drop of the control solution to the tip of the test strip.
4. When the test strip confirmation window is full, the meter will beep. The meter
will count down from five seconds and then show the reading. Make sure the
reading on the screen is between the ranges displayed on your test strip vial
(refer to page 44 for the control solution ranges).
5. Mark all control solution test with 'Control Solution Mark' to distinguish them
from blood glucose test results in the meter memory. Marked control solution
result will not be included in 14 day average. For more detailed information
of Flagging Your Test Results refer to page 22 for help.
5
42

Advertisement

Table of Contents
loading

Table of Contents