Chapter 1
FOCUS GC Description

Carrier Gas Inlet Port

It is located on the middle upper part of the instrument rear panel. Gas inlet port
are directly connected to the detector modules.
WARNING! Gas pressure must be controlled on the gauges provided on the gas tank of the
supply gas line. The maximum inlet pressure at 1000 kPa (145 psi) should be set to
allow DPFC module to adjust pressure up to 1000 kPa.
Commonly used gas are helium, hydrogen and nitrogen.
The default pressure unit is the kilopascal (kPa). You specify the pressure units in
the
The following gives a brief conversion guide for the most commonly used
pressure units.

Detector Gases Inlet Ports

They are located on the middle upper part of the instrument rear panel. Detector
gas inlet ports are directly connected to the DGFC module.
WARNING! Pressure of the supply gas lines should be set to a minimum of 400 kPa (60 psi) to
ensure the regulation of detector gases.
Maximum input pressure to FOCUS GC (from supply gas lines) is 1000 kPa (145 psi).

Servo Air for Gas Sampling Valve Inlet Port

It is located on the left upper part of the instrument rear panel. Servo air inlet port
is directly connected to the gas sampling valve when installed into the GC.
36
menu.
Configuration
To Convert
kPa
bar
psi
100 kPa = 1 bar = 14.51 psi
To
Multiply by
bar
0.01
psi
0.145
kPa
100
psi
14.51
kPa
6.89
bar
0.0689