Specifications - Honeywell IAQPoint2 Technical Manual

Indoor air quality monitor
Hide thumbs Also See for IAQPoint2:
Table of Contents

Advertisement

IAQPoint2

Specifications

Description
Gas
Detection method
Monitor with
Ranges
CO
sensor
2
Accuracy
Response Time
Calibration
Gases
Detection method
Monitor with
VOC sensor
Range
Calibration
Temperature Range
Temperature Accuracy
Humidity Range
Humidity Accuracy
Duct-mount Air Velocity 420 ft/min (2 m/s) minimum, 1760 ft/min (9 m/s) minimum
Power in
System
Consumption
Operations
Dimensions
Weight
Enclosure
Display
Languages
Storage temperature
Analog
Digital
Outputs
Relay
Wall mount
Mounting
options
Duct mount
US
Canada
Certifications
Europe
Compliance
Warranty
1 allow an additional ±2% of reading for calibration gas cylinder
2 self-calibration period: 2-3 weeks; warm-up time to spec precision: 1 minute
3 additional uncertainty of 2% of the reading for readings greater than 2,000 ppm
4 in still air after calibration
50
CO
2
Non-dispersive infrared (NDIR) with automatic background calibration
(ABC) algorithm
0-2,000 ppm, 0-5,000 ppm
±30 ppm ±3% of reading at 25°C from 0 to 2,000 ppm
T
<60 seconds
90
Not required
Volatile organic compounds (VOCs)
Metal oxide semiconductor (MOS)
0-100 air quality index
Not required
32°F to 122°F (0°C to 50°C), 32°F to 104°F (0°C to 40°C) for wall-mount
±0�5°C at 25°C and 50% rH ±1�5°C over range
0 to 95% rH, non-condensing
±5% at 25°C and 50% rH
20 to 30 VAC; 18 to 28 VDC
typical: 188 mA @ 24 VAC (150 mA minimum, 281 mA maximum)
Analog
typical: 87 mA @ 24 VDC (75 mA minimum, 174 mA maximum)
typical: 174 mA @ 24 VAC (135 mA minimum, 215 mA maximum)
Digital
typical: 83 mA @ 24 VDC (75 mA minimum, 120 mA maximum)
3�0 in� W x 4�6 in� H x 1�5 in� D (77 mm x 118 mm x 37 mm)
0�44 lbs (0�20 kg)
ABS
TFT, multicolor graphical touchscreen, 1�9 in� W x 2�6 in� H (49 mm x 65
mm); optional with analog units
English, French for initial release
to 122ºF (-10ºC to 50ºC)
4-20 mA, 0-10 VDC selectable
Modbus RTU, BACnet MS/TP
3 amps, 250 VAC resistive, and 30 VDC SPDT; field configurable set-
point and hysteresis
IP22
IP22
ANSI/UL 61010-1
CAN/CSA C22�2 No� 61010-1
CE EN 13779-2007
RoHS, WEEE
One year
Value
1, 2, 3
4
4

Advertisement

Table of Contents
loading