Net-Total Formulation R

When the Net-Total Formulation
application is selected, you can weigh
in different components up to a defined
total. Each component is saved in the
net-total memory.
You can print out both the total
weight and the individual weights of
the components.
Features:
– Weigh in up to 199 components
in series
– Current component number shown
in the numeric display (indicating the
component to be added)
– Toggle the display between
"component mode" and "additive
mode" by pressing w:
– Component mode: Display the
weight of the component currently
on the instrument (for 1 second
after it is saved; then the instrument
is tared)
– Additive mode: Display the weight
of all components on the instrument
(after it is saved, the net weight
of the last component added is
displayed briefly)
– Toggle to a second weighing instrument
while weighing on the first
– Automatic printout of component
or total weight when weight is saved
– Automatic taring of container weight
– Minimum load for automatic taring
– Minimum load for storing values
– Display the value in totalizing memory
in the Info mode, accurately calculated
in the active weight unit, by pressing
and holding w
52
Application Parameters: Net-Total Formulation
3. 5.
Minimum Load for Automatic Taring
and Automatic Printing
3. 5. 1 * 1 digit
3. 5. 2
2 digits
3. 5. 3
5 digits
3. 5. 4
10 digits
3. 5. 5
20 digits
3. 5. 6
50 digits
3. 5. 7
100 digits
3. 5. 8
200 digits
3. 5. 9
500 digits
3. 5.10
1000 digits
3. 6.
Minimum Load for Automatic
Storage/Transfer of Values
3. 6. 1 * 1 digit
3. 6. 2
2 digits
3. 6. 3
5 digits
3. 6. 4
10 digits
3. 6. 5
20 digits
3. 6. 6
50 digits
3. 6. 7
100 digits
3. 6. 8
200 digits
3. 6. 9
500 digits
3. 6.10
1000 digits
3. 7.
Automatic Taring: 1st Weight Tared
3. 7. 1 * Off
3. 7. 2
On
3.17.
Automatic Individual or Component
Printout when Value Stored
3.17. 1
Off
3.17. 2* Generate printout with complete
standard configuration each time
O is pressed
3.17. 3
Generate printout with complete
standard configuration only once
when O is pressed
Minimum Load for Automatic Storage/
Transfer of Values
The minimum load for automatic
storage or transfer of weight values is
configured under menu item 3.6.
If the load exceeds this limit, the weight
currently displayed can be stored in net-
total memory.
Net-Total Formulation with
Two Weighing Instruments
This mode is used for weighing
large and small components at the
same time.
In this mode, you can toggle from
the small-component instrument
to the large-component instrument
during measurement. Once you toggle
to the large-component instrument,
the ( and ) keys are available
until a component is value is saved.
For example, you can tare a partially-
filled container taken from the small-
component instrument on the large
component instrument.
The value in component memory on
the small-component instrument is
transferred to the large-component
instrument and the weight unit is con-
verted, if necessary. The Component
and Additive display modes are both
available on the large-component
instrument.
The value read by the active instrument
is saved in component memory. The
displayed result is accurately calculated
in the active weight unit, for example:
1243 g
(from an instrument with 1 interval)
+ 1.40 kg
(from an instrument with 5 intervals)
= 2.643 kg (calculated result)
When you press c to stop a measure-
ment series, the tare memories for
both weighing instruments are cleared,
unless the large-component instrument
is in SBI mode, in which case the
instrument is only tared.