Sartorius CH1NE Operating Instructions Manual page 26

Combics complete scales
Table of Contents

Advertisement

External calibration
Nom.
+
10000.0 g
Diff. -
0.3 g
--------------------
)
--------------------
14.01.2002
13:00
Model CW2P1-30ED-LCE
Ser.no.
12345678
Vers.
1.1007.12.1
BVers.
01-25-01
--------------------
External calibration
Nom.
+
10000.0 g
Diff. -
0.3 g
External adjustment
Diff. +
0.0 g
--------------------
14.01.2002
13:02
Name:
--------------------
26
Position the calibration weight on the weighing instrument
The difference between the weight value and the true mass is displayed, with ± sign.
Calibration record is printed, if adjustment was not performed and the process
was stopped by pressing (
Activate calibration/adjustment manually (press the ( key to cancel)
The calibration weight is displayed at the conclusion of calibration
A GMP-compliant printout is generated
Unload the weighing instrument