Alarms (Continued)
Use the s or t key to toggle between Alarm On or Off. Press the Enter key.
Use the s or t key to cycle through the following data sources: Temp C, Temp K, Linear, or Sensor,
where Temp C = degrees Celsius, Temp K = kelvin, Linear = MX+B or M(X+B) (refer to Paragraph
4.14.2), or Sensor = volts (V), millivolts (mV) or ohms (Ω). Press the Enter key.
The Low Alarm Point is entered using the numeric keypad, which includes the numbers 0 – 9, +/–, and
decimal point. For this example, enter 50 K. Press the Enter key.
The High Alarm Point is entered using the numeric keypad, which includes the numbers 0 – 9, +/–,
and decimal point. For this example, enter 100 K. Press the Enter key.
Use the s or t key to toggle between Latching On or Off. For this example, select Alarm Latching
Off. Press the Enter key.
The dead band is entered using the numeric keypad, which includes the numbers 0 – 9, +/–, and
decimal point. Press the Enter key.
The audible parameter determines whether the internal beeper will sound when an alarm is active.
This is a global parameter so it is set once for all alarms.
After specifying either Alarm Setpoint On or Dead Band, next is the Alarm Audible screen.
Use the s or t key to toggle between Audible Alarm On or Off. For this example, select Audible
Alarm On. Press the Enter key.
4-30
Lake Shore Model 332 Temperature Controller User's Manual
Select for Alm A
Alarm
On
Select for Alm A
Source
Enter for Alarm A
Alarm Low
Enter for Alarm A
Alarm High
Select for Alm A
Alarm Latching
Enter for Alarm A
Dead Band
Select With °®
Alarm Audible
°®
°®
Temp K
50½
100½
°®
Off
On
Operation

Advertisement