Lakeshore 332 User Manual page 138

Temperature controller
Table of Contents

Advertisement

TEMP?
Thermocouple Junction Temperature Query
Input:
TEMP?
Returned: <junction temperature>[term]
±nnnnnnn
Format:
Remarks: Temperature is in kelvin.
TUNEST?
Control Tuning Status Query
Input:
TUNEST?
Returned: <tuning status>[term]
n
Format:
ZONE
Control Loop Zone Table Parameter Command
Input:
ZONE <loop>, <zone>, <top value>, <P value>, <I value>,
<D value>, <mout value>, <range>[term]
n,nn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,n[term]
Format:
<loop>
<zone>
<top value>
<P value>
<I value>
<D value>
<mout value>
<range>
Remarks: Configures the control loop zone parameters. Refer to Paragraph 2.9.
Example: ZONE 1,1,25.0,10,20,0,0,2[term] – Control Loop 1 zone 1 is valid to 25.0 K with P = 10,
I = 20, D = 0, and a heater range of 2.
ZONE?
Control Loop Zone Table Parameter Query
Input:
ZONE? <loop>, <zone>[term]
Format: n,nn
<loop>
<zone>
Returned: <top value>, <P value>, <I value>, <D value>, <mout value>, <range>[term]
Format: ±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,n (Refer to command for description)
6-40
Lake Shore Model 332 Temperature Controller User's Manual
0 = no active tuning, 1 = active tuning.
Specifies which loop to configure: 1 or 2.
Specifies which zone in the table to configure. Valid entries are: 1 – 10.
Specifies the top temperature of this zone.
Specifies the P for this zone: 0.1 to 1000.
Specifies the I for this zone: 0.1 to 1000.
Specifies the D for this zone: 0 to 200%.
Specifies the manual output for this zone: 0 to 100%.
Specifies the heater range for this zone if <loop> = 1. Valid entries: 0 – 3.
Specifies which loop to query: 1 or 2.
Specifies which zone in the table to query. Valid entries: 1 – 10.