INTYPE
Input Type Parameter Command
Input:
INTYPE <input>, <sensor type>, <compensation>[term]
a,n,n
Format:
<input>
<sensor type>
<compensation> Specifies input compensation where 0 = off and 1 = on. Reversal for thermal
Remarks: Sensor type NTC RTD 75mV 7.5kΩ listed twice to maintain compatibility with Model 331
INTYPE command.
Example: INTYPE A,0,0[term] – Sets Input A sensor type to silicon diode.
INTYPE?
Input Type Parameter Query
Input:
INTYPE? <input>[term]
Format: a
<input>
Returned: <sensor type>, <compensation>[term]
Format: n,n
KEYST?
Keypad Status Query
Input:
KEYST?[term]
Returned: <keypad status>[term]
n
Format:
Remarks: Returns keypad status since the last KEYST?. KEYST? returns 1 after initial power-up.
KRDG?
Kelvin Reading Query
Input:
KRDG? <input>[term]
Format: a
<input>
Returned: <kelvin value>[term]
Format: ±nnnnnn
Remarks: Also see the RDGST? command.
Remote Operation
Lake Shore Model 332 Temperature Controller User's Manual
Specifies input to configure: A or B.
Specifies input sensor type. Valid entries:
0 = Silicon Diode
1 = GaAlAs Diode
2 = Platinum 100/250 Ω
3 = Platinum 100/500 Ω
4 = Platinum 1000 Ω
5 = NTC RTD 75mV 7.5 kΩ
6 = Thermocouple 25 mV
7 = Thermocouple 50 mV
EMF compensation if input is resistive, room compensation if input is
thermocouple. Always 0 if input is a diode.
Specifies input to query: A or B.
(Refer to command for description)
1 = key pressed, 0 = no key pressed.
Specifies which input to query: A or B.
8 = NTC RTD 75mV 75 Ω
9 = NTC RTD 75mV 750 Ω
10 = NTC RTD 75mV 7.5 kΩ
11 = NTC RTD 75mV 75 kΩ
12 = NTC RTD 75mV Auto