RAMP
Control Setpoint Ramp Parameter Command
Input:
RAMP <loop>, <off/on>, <rate value>[term]
n,n,±nnnnn
Format:
<loop>
<off/on>
<rate value> Specifies setpoint ramp rate in Kelvin per minute from 0.1 to 100. The rate is
Example: RAMP 1,1,10.5[term] – When Control Loop 1 setpoint is changed, ramp the current setpoint
to the target setpoint at 10.5 K/minute.
RAMP?
Control Setpoint Ramp Parameter Query
Input:
RAMP? <loop>
Format: n
<loop>
Returned: <off/on>, <rate value>[term]
Format: n,±nnnnn
RAMPST?
Control Setpoint Ramp Status Query
Input:
RAMPST? <loop>[term]
Format: n
<loop>
Returned: <ramp status>[term]
Format: n
<ramp status> 0 = Not ramping, 1 = Setpoint is ramping.
RANGE
Heater Range Command
Input:
RANGE <range>[term]
n
Format:
RANGE?
Heater Range Query
Input:
RANGE?[term]
Returned: <range>[term]
n
Format:
Remote Operation
Lake Shore Model 332 Temperature Controller User's Manual
Specifies which loop to configure: 1 or 2.
Specifies whether ramping is 0 = Off or 1 = On.
always positive, but will respond to ramps up or down.
Specifies which loop to query: 1 or 2.
(Refer to command for description)
Specifies which loop to query: 1 or 2.
0 = Off, 1 = Low (0.5 W), 2 = Medium (5 W), 3 = High (50 W)
(Refer to command for description)